    CS600高精度碳硫測定儀

    簡要描述:簡單說明:CS600提供了一系列*的解決方案檢測碳和硫。當今的燃燒和紅外相結合的檢測技術使您能更準確、精確、使用方便靈活的剖析各種類型的金屬和無機材料。

    • 產品型號:CS600
    • 廠商性質:生產廠家
    • 更新時間:2025-07-17
    • 訪  問  量:2485
    詳情介紹

    詳細說明:

    CS600提供了一系列*的解決方案檢測碳和硫。當今的燃燒和紅外相結合的檢測技術使您能更準確、精確、使用方便靈活的剖析各種類型的金屬和無機材料。

    Carbon and Sulfur in Steel, Nickel-Base, and Cobalt-Base Alloys
     
    Instrument
    CS600-Series
     
    Sampling and Sample Preparation
    Surface contamination on the sample can cause significant errors in the analytical data; therefore, care must be taken to ensure a clean, representative sample is analyzed. Solid
    samples should be abraded with a clean file, rinsed in acetone and dried with warm air prior to analysis. Samples that cannot be abraded due to irregular shapes should be rinsed in a
    suitable solvent such as acetone, and dried with warm air. Care
    must be taken to remove all traces of the solvent. If a sample is porous, refrain from using solvents,
    as it will be difficult to remove all traces of the solvent by drying. Refer to ASTM E1806 for additional sampling and sample preparation information.
     
    Accessories
    528-018 or 528-018HP Ceramic Crucibles (preheated)*; LECOCEL II (501-008) or
    LECOCEL II HP (502-173) accelerator.
     
    *Ceramic crucibles are baked in a muffle or tube furnace (LECO TF-10) at 1250oC for a minimum of
    15 minutes, or at 1000oC for 40 minutes. The crucibles are removed from the furnace, allowed to cool for

     
     
     
     
    1 to 2 minutes, and are transferred to a desiccator for storage. If the crucibles are not used within four hours, they should be re-baked.

     
    Calibration
    There are several suitable calibration samples available from LECO. Likewise, NIST, JK, JSS, and BCS are certified bodies that have a variety of certified reference materials (SRM/CRM) available as well. It is important that both high- and low-carbon ranges are calibrated. Single or multipoint calibration curves
    can be utilized. Refer to the operator's instruction manual for details.
     
    Method Parameters
    Purge Time (seconds):
    1**
     
    Delay Time (seconds):
    8**
     
    Furnace Low Power (%):
    100
     
    Furnace High Power (%):
    100
     
    Furnace Ramp Rate:
    0
     
     
     
    Carbon
     
    Sulfur
    Minimum Timeout (seconds):
    40
    40
    Comparator Level:
    1.00
    1.00
    Significant Digits:
    4 or 5
    4 or 5
    Integration Delay:
    0
    0
     
     


     

    **For optimum analytical performance on samples with low carbon and
    sulfur content 0.05%), a purge time of 10 seconds and a delay time of
    20 seconds is recommended.

     
     
    CS600


     

    Procedure
    1.   Prepare instrument for operation as outlined in the operator's instruction manual.
    2.   Determine blank.
    a.     Enter 1.0000 g weight into weight stack. b.      Add ~1.5 g of accelerator to crucible.
    c.      Place the crucible on the furnace pedestal (or appropriate autoloader position if so equipped), and analyze.
    d.     Repeat steps 2a through 2c a minimum of five times.
    e.     Enter blank following procedure outlined in the operator's instruction manual.
    3.   Calibrate.
    a.     Weigh ~1.0 g calibration sample into crucible and enter weight into weight stack.
    b.     Add ~1.5 g of accelerator on top of sample.
    c.      Place the crucible on the furnace pedestal (or appropriate autoloader position if so equipped), and analyze.
    d.     Repeat steps 3a through 3c a minimum of five times for each calibration sample intended for calibration.
    e.     Calibrate using the procedure outlined in the operator's instruction manual.
    4.   Analyze Samples.
    a.     Weigh ~1.0 g sample into crucible and enter weight into weight stack.
    b.     Add ~1.5 g of accelerator on top of sample.
    c.      Place crucible on furnace pedestal (or appropriate autoloader position if so equipped), and analyze.
     
    Note:   Some "solid" samples, notably nickel and/or cobalt-base alloys, are difficult to combust and may require additional accelerator. Likewise, reducing the sample to small chunks, chips, or pieces will facilitate combustion. For additional information on Carbon and Sulfur Determination using an Induction Furnace IR detection instrument, request LECO's Application Tips CD (209-050-055).
     
    Typical Results

    Sample
    Weight (g)
    C %
    S %
    Sample
    Weight (g)
    C %
    S %
    LECO
    1.0
    0.605
    0.0176
    LECO
    1.0
    0.00089
    0.00062
    501-505
     
    0.605
    0.0172
    502-348
     
    0.00089
    0.00067
    Steel Ring
     
    0.605
    0.0169
    Low C and S
     
    0.00094
    0.00068
    0.607 % C
     
    0.607
    0.0171
    Steel Pin
     
    0.00095
    0.00060
    0.0172% S
     
    0.607
    0.0174
    0.0009 % C
     
    0.00085
    0.00064
     
     
    0.605
    0.0176
    0.0007 % S
     
    0.00085
    0.00061
     
     
    0.606
    0.0178
     
     
    0.00090
    0.00062
     
     
    0.604
    0.0173
     
     
    0.00090
    0.00060
     
     
    0.606
    0.0173
     
     
    0.00090
    0.00065
     
     
    0.605
    0.0168
     
     
    0.00089
    0.00055
     
    X =
    0.606
    0.0173
     
    X =
    0.00090
    0.00062
     
    s =
    0.001
    0.0003
     
    s =
    0.00003
    0.00004
     
    LECO
     
    1.0
     
    0.0375
     
    0.0223
     
    LECO
     
    1.0
     
    0.0414
     
    0.0025
    501-501
     
    0.0380
    0.0221
    501-933
     
    0.0414
    0.0027
    Steel Ring
     
    0.0375
    0.0222
    Ni-Base Alloy
     
    0.0413
    0.0027
    0.0377 % C
     
    0.0377
    0.0222
    Chip Sample
     
    0.0414
    0.0028
    0.0220 % S
     
    0.0376
    0.0223
    0.042 % C
     
    0.0415
    0.0027
     
     
    0.0376
    0.0222
    0.003 % S
     
    0.0413
    0.0028
     
     
    0.0373
    0.0221
     
     
    0.0414
    0.0028
     
     
    0.0376
    0.0222
     
     
    0.0415
    0.0028
     
     
    0.0377
    0.0223
     
     
    0.0412
    0.0027
     
     
    0.0377
    0.0222
     
     
    0.0414
    0.0029
     
    X =
    0.0376
    0.0222
     
    X =
    0.0414
    0.0027
     
    s =
    0.0002
    0.0001
     
    s =
    0.0001
    0.0001


     